General
Power supplies for laboratory use are mostly only power sources. They are only able to deliver power with one single polarity, i.e. these instruments only work in one quadrant of the output voltage/ current diagram. Common amplifiers deliver voltages with both priorities, but are generally unable to absorb power values in a range comparable to that which they are able to deliver. They therefore mainly work as two-quadrant devices. TOE 7610 series of four-quadrant amplifiers is able to work as a power sink as well as a power source with equal values of delivered or absorbed power.
The TOE 7610 series is thus a voltage amplifier, bipolar voltage and current source, and current sink in one instrument. Having to combine different instruments, where problems frequently occur with the coordination of individual components and the tendency to oscillate, now belongs to the past.
Power Manager
A selectable Power Manager enables the maximum output power Pmax. to also be processed as a continuous sink power. Numerous setting options of the 4-quadrant amplifier allow optimum connection of the amplifier with the seriesconnected driving equipment, whether this is a laboratory power supply, a calibrator, a function generator or a signal source from a PC in the form of a D/A converter card.
Selectable input impedance
The input impedance can be switched between 50 Ohm and 100 kOhm and allows convenient selection of a function generator to be controlled and a signal source in the form of a D/A card in the PC without the need for external adaptation.
Selectable input voltage range
The adjustable input voltage range allows switching between an input voltage of ±5 V or ±10 V. Therefore no further external adaptation of the control voltage is necessary.
Variable DC offset
Brief load current
All 4-quadrant amplifiers of the TOE 7610 series can be equipped with the TOE 7610/103 option. They are then able to deliver three times the rated current for approx. 3 ms.
Output ON/OFF
A convenient detail of the functionality is the output switch-off which allows immediate switching-off of the voltage and current when in the “Output OFF“ position.
Special features
Model | Output voltage | Output current | Output power |
TOE 7610-10 TOE 7610-20 TOE 7610-40 TOE 7610-60 |
± 10 V ± 20 V ± 40 V ± 60 V |
± 15 A ± 7,5 A ± 4 A ± 2,5 A |
150 W 150 W 160 W 150 W |
Rise and fall times of TOE 7610-60 |
Small signal bandwidth of TOE 7610 |
TOE 7610-10 | TOE 7610-20 | TOE 7610-40 | TOE 7610-60 | |
Output voltage Output current Current limiting (Slow mode only) CURR LIMIT POS CURR LIMIT NEG |
± 10 V ± 15 A 0 - +15 A 0 - -15 A |
± 20 V ± 7.5 A 0 - +7.5 A 0 - -7.5 A |
± 40 V ± 4 A 0 - +4 A 0 - -4 A |
± 60 V ± 2.5 A 0 - +2.5 A 0 - -2.5 A |
Power (continuously at Imax) Sink mode with Sink mode with |
150 W 150 W |
160 W 160 W |
150 W 150 W |
|
Frequency range DC input coupling Speed slow Speed fast AC input coupling Speed slow Speed fast |
0 Hz - 30 kHz 0 Hz - 100 kHz 2 Hz - 30 kHz 2 Hz - 100 kHz |
|||
Amplifier Input V Range ± 5 V var gain Off var gain On Input V Range ± 10 V var gain Off var gain On |
6 dB 6 dB ± 1 dB 0 dB 0 dB ± 1 dB |
12 dB 12 dB ± 1 dB 6 dB 6 dB ± 1 dB |
18 dB 18 dB ± 1 dB 12 dB 12 dB ± 1 dB |
21.6 dB 21.6 dB ± 1 dB 15.6 dB 15.6 dB ± 1 dB |
Input impedance DC input coupling Input Impedance 50 Ω Input Impedance 100 kΩ AC input coupling Input Impedance 50 Ω Input Impedance 100 kΩ |
50 Ω 100 kΩ 50 Ω 100 kΩ |
|||
Max. input voltage (continuous) | 12 Veff | |||
Rise time (fast mode only) | Approx. 6 V/μs | Approx. 10 V/μs | Approx. 16 V/μs | Approx. 25 V/μs |
Noise Up to 1 kHz Up to 10 MHz |
< 0.1mVrms < 1mVrms |
|||
Harmonic distortion Speed Slow Up to 1 kHz Up to 7 kHz Up to 10 kHz Speed Fast Up to 20 kHz Up to 40 kHz Up to 60 kHz |
< 0.25 % < 0.80 % < 1.00 % < 0.20 % < 0.30 % < 0.50 % |
< 0.10 % < 0.30 % < 0.50 % < 0.10 % < 0.30 % < 0.50 % |
Amplifier output | Floating, insulation ± 260 V against ground |
Line voltage | 115 V /230 V ± 10 %, 47 Hz to 63 Hz |
Line fuse | 115 V: T4L 230 V: T2L in accordance with IEC 127-2/III, DIN 41662 |
Power consumption | Max. approx. 400 VA |
Protective measures | Protection class I in accordance with DIN EN 61010-1 |
Operating temperature | 0 °C to 40 °C |
Storage temperature | 0 °C to 40 °C |
Reference temperature | -20 °C to 70 °C |
Cooling | With thermostatically controlled fan |
Dimensions (W x H x D) With feet |
216 x 132 x 429 mm 216 x 147 x 429 mm |
19“ system | Compatible with ½ 19“, 3 HU |
Weight | Approx. 9 kg |
Housing | Aluminium |
TOE 7610-10 | DC - 100 kHz, ± 10 V / ± 15 A |
TOE 7610-20 | DC - 100 kHz, ± 20 V / ± 7,5 A |
TOE 7610-40 | DC - 100 kHz, ± 40 V / ± 4 A |
TOE 7610-60 | DC - 100 kHz, ± 60 V / ± 2,5 A |
TOE 9508 | Carrying handle |
TOE 9502 | 19“ adapter, 3 HU for single installation |
TOE 9508 | 19“ adapter, 3 HU, parallel installation set for 2 units |
TOE 7610/103 | Brief load current 3 x Irated for approx 3ms |
Supplied accessories